Introduction
Implantable cardioverter defibrillators (ICDs) are electronic devices that are intricately designed with the purpose of detecting as well as terminating cardiac arrhythmias. With added capabilities of demand pacing, their applications are two-fold – emergency defibrillation and backup pacing. Initially looked upon as a remote choice for patients with sustained ventricular tachycardia (VT), their effectiveness is unquestioned today [1–3]. Over 200,000 ICDs are implanted worldwide every year, with rapidly expanding applications in modern day medicine [4].
Prologue to the birth of ICDs
Beck described open chest defibrillation of the human heart for the first time in 1947 [3, 5]. A decade later, Kouwenhoven showed that it was possible to defibrillate the heart through a closed chest. The weight of the first external defibrillator so employed was a massive 120 kg and delivering an alternating current (AC) of 500 V, thus limiting its practical application [3, 6]. Lown, in 1962, developed a direct current (DC) defibrillator, which consisted of a capacitor for storing energy, a battery and a transformer to overcome the high incidence of cardiac arrhythmias and cardiac damage that occurred with AC current [3, 7]. It was not until 1970 that the use of an implantable defibrillator was first demonstrated by John Schuder, at the University of Missouri in 1970, on a canine model [2, 3, 7–9].
Conception of ICDs in humans and the early design
1980 was the landmark year that saw the first successful implantation of ICD in a human by Michel Mirowski [3, 10]. The twist of events leading to this feat was rather interesting. It was the sudden and unfortunate death of his colleague, Dr. Harry Heller in 1966, due to VT, that led Mirowski to conceive the idea of an ICD. In 1968, Mirowski was hired by the Sinai Hospital in Baltimore as the director of the coronary care unit, with ample time for research. The biomedical engineering division and animal laboratory of the hospital aided research [3, 11]. The early fruit was the successful testing of the first crude prototype on a dog in August 1969 [3, 9, 10, 12]. Eventually, Mirowski and Mower garnered support from a major pacemaker company in 1970 to further develop the ICD, only to be abandoned 2 years later, as the company saw no market for the device.
In 1972, Mirowski came in touch with Stephen Heilman, a physician-cum-engineer who founded Medrad, a medical equipment company. Heilman, who was highly impressed by the novel idea of ICD, put his engineers at Mirowski’s disposal. This partnership yielded the first ICD prototype small enough to be implanted in a dog in 1975. A film of the first successful defibrillation of a dog implanted with the prototype ICD was released and this famous footage propelled ICD into the limelight overnight [3, 11].
Further refinements in the prototype were made to make it suitable for human implantation. Food and Drug Administration (FDA) approval was eventually received. Ably supported by his colleagues Myron Weisfeldt (cardiac surgeon) and Philip Reed (electro physiologist), at Johns Hopkins Hospital, the first human implant of an ICD was performed successfully in February 1980 [3, 9, 10, 13]. A 57-year-old woman with severe coronary artery disease with recurrent VA refractory to all standard antiarrhythmic drugs was the first to receive the device [3]. The ICD could successfully terminate a ventricular flutter induced during an electrophysiological study 2 weeks later. After five months, the patient had been without VA. The second patient was a 16-year-old boy with refractory VA. Five episodes of hypotensive VT were cardioverted successfully by the device 2 weeks after implantation [3, 10].
Early design
The devices implanted in the 1980s were designed merely to recognize and terminate ventricular fibrillation (VF) by delivering a high-energy shock [10]. These early devices lacked the ability to detect unstable VT that could degenerate into VF. Moreover, as these devices were not programmable, separate pacemakers were needed to allow backup bradycardia pacing, leading to lethal interactions [10, 12, 14]. It weighed 250 g and had a volume of 145 ml. An open thoracotomy was required for implantation of the epicardial patch electrodes and the device was implanted in the abdomen. The electrodes were made of titanium and silicone rubber. One electrode was placed epicardially on the cardiac apex while a second electrode sat in the right ventricular apex. The battery was made of titanium and was hermetically sealed to prevent moisture from reaching the capacitor and other circuitry. Within 15 s of induced VF, the device delivered a 30 joule shock to convert the rhythm. The life of the lithium battery was projected to be 3 years or 100 shocks [3]. In the following years, numerous advances in ICD design were made. Mirowski and Mower pioneered several of these early refinements [3, 10, 15].
Evolution to the present day form – through the generations
As a major improvement over the early devices, the second generation ICDs were designed to detect ventricular arrhythmias (VA) using a probability density parameter based on the concept that, unlike sinus rhythm, ventricular fibrillation did not maintain an isoelectric baseline. This enabled bradycardia pacing ability and they were minimally programmable. This ended the need for separate pacemakers. The second generation ICDs could do away with thoracotomy by the introduction of transvenous leads in 1988, which enabled the implantation procedure to be performed in an electrophysiology laboratory rather than open surgery [16]. Furthermore, these devices possessed limited telemetry function to test battery strength, for which an external monitoring device was needed. Cylindrical aluminium electrolytic capacitors and silver vanadium pentoxide batteries of the first generation ICDs were replaced by lithium-silver vanadium manganese oxide batteries, which resulted in longer life of ICDs [14].
The first third generation ICDs were introduced in the early 1990s. Anti-tachycardia pacing (ATP), low energy shocks for terminating VTs, high level of programmability and telemetry functions were the key upgrades [17]. To improve specificity in discriminating between VT or supraventricular tachycardia, various algorithms were developed. ICDs could be programmed into three different cycle length-related zones and the discriminative detection algorithms programmable in the two lowest zones. The highest programmable zone is meant for detection of fast VT or VF without any further discrimination to avoid unnecessary delay of delivery of therapy. Additionally, improvements were made in lead construction. The coaxial lead design of the first and second generation ICDs was replaced with the multi-lumen lead design in third generation ICDs. The coaxial lead had a layered design comprising a tip conductor, ring conductor, defibrillation conductor and an insulation layer between each conductor. The multi-lumen lead construction is based on parallel running conductors through a single insulating body. The key advantage of multi-lumen over coaxial leads is the fact that more conductors would fit into overall smaller leads. The tip and ring conductors are used for pacing and sensing, a defibrillation conductor for the coil located in the right ventricle and a defibrillation conductor for the coil located in the superior vena cava. The insulating body contains extra lumens to increase the lead’s resistance to compression forces [18].
ICDs of today or the fourth generation ICDs have progressively become smaller and more sophisticated. They weigh not more than 80–90 g and have a volume of 30 ml, measuring less than a centimeter in thickness. Newer lithium silver vanadium batteries now last up to 9 years [19]. All modern ICDs carry the ability of overdrive pacing, i.e. ATP, which can often terminate VTs without resorting to shock therapy. In addition, ICDs are also available with biventricular pacing (cardiac resynchronization therapy) to improve symptoms in patients with advanced cardiac failure [2].
Indications for ICDs
Indications for ICDs can be broadly categorized as primary prophylaxis and secondary prophylaxis against SCD. Multiple randomized clinical trials have shown ICDs to be clearly superior to antiarrhythmic drugs in patients with a history of life-threatening VT and VF and therefore the indications for secondary prophylaxis are well supported by clinical evidence. However, indications for primary prophylaxis, representing most of the ICD implants, have comparatively less well established evidence, as the measurable quantitative benefit is smaller [1].
Wilcox et al. [20] studied the clinical effectiveness of cardiac resynchronization (CRT) and ICD therapy among 8936 outpatients with heart failure and reduced ejection fraction (EF) ≤ 35% and found a substantially reduced 24-month mortality in eligible men and women with heart failure and reduced EF. Pokorney et al. [21] investigated the relationship between primary prophylaxis of ICDs and mortality in 852 racial/ethnic minority patients (non-white race or Hispanic ethnicity) and 2070 white non-Hispanic patients. Minority ICD patients showed a lower adjusted 3-year mortality rate (44.9%) as compared to their non-ICD counterparts (54.3%). Similarly, white non-Hispanic ICD patients showed a lower adjusted 3-year mortality rate (47.8%) in comparison to their non-ICD counterparts (57.3%). No significant relationship between race/ethnicity and lower mortality risk with ICD was found and the results concur with the use of a similar approach in selecting ICD patients, irrespective of race or ethnicity.
Several randomized control trials have shaped the present day indications for ICDs. Some of the major trials include MADIT I & II, CAT, MADIT-CRT, PRESERVE EF and PROTECT-ICD. The MADIT-I study showed that ICD saves lives in high-risk patients with coronary heart disease. The MADIT-II study showed that prophylactic ICD therapy was associated with significantly improved survival in patients with ischemic cardiomyopathy, as defined by documented coronary heart disease and advanced left ventricular dysfunction, without requiring screening for ventricular arrhythmias or inducibility by electrophysiologic testing. Together, these studies, as well as the results from several other randomized ICD trials, indicate that ICD therapy is indicated in coronary patients who meet MADIT-I or MADIT-II eligibility criteria, who are not excluded by major non-cardiac comorbidity [22]. The CAT trial did not provide evidence in favor of prophylactic ICD implantation in patients with DCM of recent onset and impaired left ventricular ejection fraction [23]. Patients with New York Heart Association class I or II heart failure, ejection fraction ≤ 30% and left bundle branch block (LBBB) derive substantial clinical benefit from CRT-D in the form of a reduction in heart failure progression and a reduction in the risk of ventricular tachyarrhythmias. Cardiac resynchronization therapy (CRT) thus evolved as a Class I treatment indication with Level of Evidence A, in patients with mild heart failure, depressed left ventricular ejection fraction, and wide QRS [24]. Current classification of SCD vs. NSCD fails to identify VTA events at the time of death in a significant proportion of patients. Multivariate analysis showed that factors independently associated with VTA mortality included VT/VF > 72 h prior to the mortality event, hospitalization for heart failure, and a history of hypertension. SCD annual incidence is 0.6–1% in post-MI patients with left ventricular ejection fraction (LVEF) ≥ 40%. No recommendations for ICD exist in this population. The two-step approach of the PRESERVE EF study detects a subpopulation of post-myocardial infarct (post-MI) patients with preserved LVEF at risk for MAEs that can be effectively addressed with an ICD [25]. The Programmed Ventricular Stimulation to Risk Stratify for Early Cardioverter-Defibrillator Implantation to Prevent Tachyarrhythmias following Acute Myocardial Infarction (PROTECT-ICD) trial is an ongoing trial targeting the prevention of sudden cardiac death in patients who have at least moderately reduced cardiac function following a MI. The primary objective of the trial is to assess whether electrophysiological study to guide prophylactic implantation of an ICD early following MI (first 40 days) will lead to a significant reduction in sudden cardiac death and non-fatal arrhythmia. The secondary objective is to assess the utility of cardiac MRI (CMR) in assessing early myocardial characteristics, and its predictive value for both inducible ventricular tachycardia (VT) at EPS and SCD/non-fatal arrhythmia at follow-up [26].
The complete list of indications for ICDs are listed in Table I as per the 2017 American College of Cardiology, American Heart Association and Heart Rhythm Society (ACC/AHA/HRS) guidelines [1].
Table I
Indications [1] | Class [1] |
---|---|
| Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] |
| Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] |
| Class IIb (The benefit probably outweighs the risk and the treatment may be considered) [1] Class IIb (The benefit probably outweighs the risk and the treatment may be considered) [1] |
| Class III (The risk outweighs the benefit, and the treatment should not be performed) [1] |
[i] IHD – ischemic heart disease, VT/VF – ventricular tachycardia/ventricular fibrillation, LVEF – left ventricular ejection fraction, SCA – sudden cardiac arrest, EPS – electrophysiological studies, GDMT – guideline-directed management and therapy, NYHA – New York Heart Association, HF – heart failure, NICM – non-ischemic cardiomyopathy, LVAD – left ventricular assist device, CRT – cardiac resynchronization therapy, HOCM – hypertrophic obstructive cardiomyopathy, NSVT – non-sustained ventricular tachycardia.
Wearable cardioverter defibrillators (WCDs), subcutaneous implantable cardioverter defibrillators (S-ICDs) and future perspectives
Introduced in 1998 [27], at the University Hospital Magdeburg, Germany, wearable cardioverter defibrillators (WCDs) were approved by the US FDA in 2001. WCDs are now frequently recommended for patients considered to be at high arrhythmic risk with potentially reversible underlying pathology or who are awaiting an ICD implantation to bridge a period of high risk of SCD [28].
A well-fitting electrode belt is necessary to ensure reliable arrhythmia detection. The harness comprises four sensing electrodes (anterior, posterior, right and left) that generate a two-lead filtered ECG. The arrhythmia detection algorithm combines heart rate data and morphology analysis. The patient’s consciousness is evaluated by assessing the response to the escalating alerts. Response buttons are present on either side of the monitor. The patient is required to press and hold both buttons simultaneously for therapy to be withheld. If the patient does not press or release them, the escalating warning system continues until an audible message to stand clear is emitted for bystanders. The patient interaction feature contributes to optimization of the sensitivity and specificity of the algorithm [29]. Patient compliance is mandatory for appropriate performance of the WCD. Careful training and instruction prior to WCD wearing are very important. Appropriate patient training for the correct use of the WCD is provided by the technical personnel of the manufacturer and includes instructions on battery charging and data transfer to the network server. After the initial training, which may last an hour or more, the technical personnel contact the patient within the next few days to avoid potential problems of device use. Patients or their relatives can contact the 24/7 hotline for queries or in case of problems.
The ACC/AHA/HRS Guidelines 2017 included WCDs. The indications for the use of WCDs are listed in Table II [1]. WCDs have been successful not only in terminating arrhythmic events, but also in preventing unnecessary ICD implantation and its associated morbidity.
Table II
Indication [1] | Class [1] |
---|---|
In patients with ICD and a history of SCA or sustained VA in whom removal of ICD is required (e.g. infection) [1] | Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] |
In patients at increased risk of SCD but not ineligible for ICD, such as awaiting cardiac transplant, LVEF ≤ 35% and are < 40 days post-MI, or have newly diagnosed NICM, revascularization within the past 90 days, myocarditis or secondary cardiomyopathy [1] | Class IIb (The benefit probably outweighs the risk and the treatment may be considered) [1] |
An entirely subcutaneous ICD (S-ICD) is a significant step forward towards a less invasive approach. As transvenous ICDs still carry considerable morbidity (~1.5% major complications) [30] and other procedural risks such as endocarditis, vessel occlusion, lead dislodgment, valvular dysfunction and intrinsic lead defects/failure and infections, S-ICDs offer an attractive alternative to prevent these shortcomings of transvenous leads. The S-ICDs received the CE mark and US FDA approval in June 2009 and September 2012, respectively [31].
S-ICDs are equipped with extrathoracic, subcutaneous electrodes, thus completely avoiding transvenous leads and related morbidity. The defibrillation coil (8 cm in length) lies directly between two sensing electrodes, while the S-ICD generator acts as the 3rd electrode, used for sensing and defibrillation. The pulse generator serves as a mandatory component of the defibrillation pathway and as an optional electrode for sensing [31]. The newest generation of S-ICD incorporates wireless connectivity to secure servers to deliver health-related information to remote caregivers. S-ICDs require a careful selection process to avoid both oversensing (T-wave) and undersensing (R-wave) leading to catastrophic consequences. One of their potential limitations, however, is in cases where a higher defibrillation threshold is to be expected (e.g. HCM) [32].
The 2017ACC/AHA/SRS guidelines included the use of S-ICD. The current indications for the use of S-ICD are listed in Table III [1].
Table III
Indication | Class |
---|---|
In patients meeting the criteria for ICD with inadequate vascular access or at high risk for infection, and in whom pacing for bradycardia or VT termination or as part of CRT is neither needed nor anticipated [1] | Class I (The treatment is useful, its benefit greatly outweighs the risk, and it should be administered) [1] |
In patients meeting the indication for ICD, if pacing for bradycardia or VT termination or as part of CRT is neither needed nor anticipated [1] | Class IIa (The benefit outweighs the risk and it is reasonable to administer the treatment) [1] |
In patients meeting the indication for bradycardia pacing or CRT, or for whom ATP for VT termination is required, S-ICD should not be implanted [1] | Class III (The risk outweighs the benefit, and the treatment should not be performed) [1] |
Future perspectives
ICDs will continue to evolve, become smaller, equipped with more advanced detection algorithms to treat life-threatening arrhythmias. In the future, the potential integration of S-ICDs with leadless pacing may make this therapy suitable for a larger target population at high risk of SCD. Moreover, if the leadless pacemaker is epicardial, left atrial and left ventricular pacing can be introduced and linked to S-ICD; alternatively application of a left ventricular seed would permit resynchronization [33]. S-ICDs may also serve as a remote monitoring device to recognize atrial fibrillation, acute ischemia, or electrolyte imbalance [34].
New ICD and CRT-D devices such as the new Gallant (Abbott Inc., USA) have received the CE Mark recently in February 2020. This new Gallant system, for instance, pairs with the secure myMerlinPulse mobile app to help streamline communication and increase engagement between doctors and their patients. This sort of app-based ICD system which would allow patients to engage more frequently with their healthcare team by providing access to transmission history and device performance and schedule their appointments could be the immediate future of ICDs.
Application and refinements of global positioning systems (GPS) may allow remote monitoring and programming of pacemakers wherever patients are on the planet. Furthermore, these developments are ideal in rural and inaccessible areas. Incorporation of advanced technologies such as artificial intelligence (AI) in CRM devices will further help patients monitor their cardiac health by notifying them and their physicians in case of abnormal activity. These cutting edge technologies come with cybersecurity threats and would therefore require the latest cybersecurity controls and partnering with industry cybersecurity experts to provide input into the design and testing of these controls to provide a seamless and a secure experience.
Conclusions
ICDs have come a long way since their first implantation in humans in 1980, firmly cementing their place in preventing SCD. Despite four decades of continual design improvements and upgrades, ICDs continue to evolve and current research is ongoing to refine the ever expanding indications for their use. With time, it is hoped that these devices will become more cost effective and safer.